Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
Procedure
1Connect the output of the HP 3325B to Channel 1 of the Counter as shown in Figure
HP 3325B
HP 53181ASynthesizer
Counter
1
Channel 1 | Output |
Figure 1-11. Trigger Level Test Setup
2Set the HP 3325B to output a 1 MHz, 80
3Set the HP 3325B DC OFFSET to −60 mV.
4On the Counter, press Run key.
5Now, increment the dc offset on the HP 3325B by +1 mV until the Counter’s Gate annunciator flashes and continue incrementing until the Counter displays approximately 1 MHz.
Observe the offset value on the display of the HP 3325B. Record the dc offset value__________ mV.
6Add the upper peak voltage (40 mV) of the 80 mV
This is the upper hysteresis level.
Record the upper hysteresis level__________ mV.