Chapter 1 Performance Tests

Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP 53181A Only)

3

Record the actual reading in the Performance Test Record

 

 

(Test 5, Line 1).

 

4

Change the synthesized signal generator’s frequency to 1.5 GHz.

1

 

 

 

Verify the Counter reads 1.5 GHz ±1 Hz

 

5

Record the actual reading in the Performance Test (Test 5, Line 2).

 

 

 

 

6

Disconnect the test setup.

 

Minimum to Maximum Accuracy Test for Options

 

030/050/124

 

Equipment

HP 8663A Synthesized Signal Generator (not needed for Option 050) HP 8340B Sweep Oscillator

HP 1250-0780 Type N-to-BNC Connector (not needed for Option 050) HP 10503A BNC Cable (2)

HP 10100C 50Ω Feedthrough

HP 11500D N-N Cable (Option 050 only) HP 1250-1250 N(M) to SMA(F)

HP 1250-0777 N(F) to N(F)

Counter Setup

1Cycle the POWER key to preset the Counter.

2Set the Counter as follows:

MEASURE/Freq & Ratio, FREQUENCY 2

Gate & ExtArm, TIME: TIME 1.000 s

Assembly-Level Service Guide

1-65