Chapter 7 HP 53131A/132A Specifications

Instrument Inputs (Continued)

Instrument Inputs (Continued)

Channel 3 Input Specifications4, 5

Frequency Range

 

 

Option 030

100

MHz to 3 GHz

Option 050

200MHz to 5 GHz

Option 124

200

MHz to 12.4 GHz

Channel 3 Input Characteristics

Impedance:

50 Ω

Coupling:

AC

VSWR:

< 2.5:1

Power Range and Sensitivity (Sinusoid)

Option 030

 

100 MHz to 2.7 GHz:

–27 dBm to +19 dBm

2.7 GHz to 3 GHz:

–21 dBm to +13 dBm

Option 050

 

200MHz to 5 GHz:

–23 dBm to +13 dBm

Option 124

 

200 MHz to 12.4 GHz

–23 dBm to +13 dBm

Damage Level:

 

Option 030

5 Vrms

Option 050

+25 dBm

Option 124

+25 dBm

7

External Arm Input Specifications

Signal Input Range:

TTL compatible

 

Timing Restrictions:

 

Pulse Width:

> 50 ns

Transition Time:

< 250 ns

Start-to-Stop Time:

> 50 ns

Damage Level:

10 Vrms

External Arm Input Characteristics

Impedance:

1 kΩ

Input Capacitance:

17 pF

Start Slope:

Positive or Negative

Stop Slope:

Positive or Negative

Notes:

External Arm is available for all measurements except Peak Volts.

External Arm is referred to as External Gate for some measurements.

4Channel 3 is available as an option.

5When ordered with optional rear terminals, the Channel 3 connector on the front panel for Option 030 will be routed to the rear panel (rear panel only). There is no degradation in specifications for this input. Option 050 and Option 124 input connectors are available on the front

panel only.

.

Assembly-Level Service Guide

7-3

Page 221
Image 221