Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
7In the Counter’s Channel 1 Trigger/Sensitivity menu, change the SLOPE to NEG.
8On the Counter, press Run key.
9Set the HP 3325B DC OFFSET to +60 mV.
10Now, decrement the offset on the HP 3325B by −1 mV until the
Counter’s Gate annunciator flashes and continue decrementing until the Counter displays approximately 1 MHz.
Observe the offset value on the display of the HP 3325B. Record the value__________ mV.
11Add the lower peak voltage (−40mV) of the 80
This result is the lower hysteresis level.
Record the lower hysteresis level __________ mV.
12Now, add the upper hysteresis value (recorded in step 6 )and lower hysteresis value (recorded in step 11).
This is the trigger level; it should be 0.0 V ±15 mV.
13Record the trigger level value in the Performance Test Record (Test 1, Line 1).