Chapter 2 Service
The HP 53181A Calibration Procedures
| The HP 53181A Calibration Procedures |
| First Determine the Counter Firmware Revision |
| Press and hold Recall (Utility) key, then cycle POWER key. |
| REV: n is displayed; where “n” represents the numeric firmware revision |
| code. |
|
|
NOTE | After completing the instrument calibration, BE SURE to secure the |
| calibration by performing the procedure in the section titled “To Secure |
| Against Calibration” on page |
|
|
To Calibrate the Offset for Channel 1
1To access the Calibration menu, power down, and press and hold the Scale & Offset key, then press POWER key.
The first item in the Calibration menu is a message indicating the calibration security status: CAL SECURE or CAL UNSECURE.
The Counter calibration must be unsecured to perform calibration.
If CAL UNSECURE is displayed, the Counter is already unsecured for calibration; go to step 5. However, if CAL SECURE is displayed then perform steps 2 through 4 to unsecure calibration.
2Press Scale & Offset key until CODE: 0 is displayed.
3Press the appropriate arrow keys to enter the security code
“53181”, for example, if 53181 is the security code by performing the following steps:
a.Press
key four times.
The Counter should display CODE: 00000.
b.Press
key five times.
The Counter should display CODE: 50000.
2