HP manual HP 53181A Calibration Procedures, To Calibrate the Offset for Channel

Models: 181A

1 254
Download 254 pages 46.13 Kb
Page 127
Image 127

Chapter 2 Service

The HP 53181A Calibration Procedures

 

The HP 53181A Calibration Procedures

 

First Determine the Counter Firmware Revision

 

Press and hold Recall (Utility) key, then cycle POWER key.

 

REV: n is displayed; where “n” represents the numeric firmware revision

 

code.

 

 

NOTE

After completing the instrument calibration, BE SURE to secure the

 

calibration by performing the procedure in the section titled “To Secure

 

Against Calibration” on page 2-29 of this chapter.

 

 

To Calibrate the Offset for Channel 1

1To access the Calibration menu, power down, and press and hold the Scale & Offset key, then press POWER key.

The first item in the Calibration menu is a message indicating the calibration security status: CAL SECURE or CAL UNSECURE.

The Counter calibration must be unsecured to perform calibration.

If CAL UNSECURE is displayed, the Counter is already unsecured for calibration; go to step 5. However, if CAL SECURE is displayed then perform steps 2 through 4 to unsecure calibration.

2Press Scale & Offset key until CODE: 0 is displayed.

3Press the appropriate arrow keys to enter the security code

53181”, for example, if 53181 is the security code by performing the following steps:

a.Press key four times.

The Counter should display CODE: 00000.

b.Press key five times.

The Counter should display CODE: 50000.

Assembly-Level Service Guide

2-31

2

Page 127
Image 127
HP manual HP 53181A Calibration Procedures, To Calibrate the Offset for Channel