Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
3Press Channel 1 50Ω/1MΩ key.
The LED turns on, indicating a 50Ω input impedance for Channel 1.
4Press Channel 1 DC/AC key.
The LED turns on, indicating DC input coupling for Channel 1.
5Verify the DVM reads 51.1Ω ± 3Ω.
6For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
6’ For HP 53181A, mark Pass or Fail in the HP 53181A Performance Test Record on page
7Press Channel 1 50Ω/1MΩ key.
The LED turns off, indicating a 1 MΩ input impedance for Channel 1.
8Verify the DVM reads approximately 1.0000000 MOHM.
9For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
9’ For HP 53181A, mark Pass or Fail in the HP 53181A Performance Test Record on page
10Repeat steps 3 through 9 for Channel 2 for the HP 53131A/132A Counter ONLY since the HP 53181A Counter does not have a standard Channel 2.
11For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
12Disconnect the test setup.
1