Chapter 2 Service

About the HP 53131A/132A Calibration Menu

 

To Secure Against Calibration

1

Press and hold Scale & Offset key, then cycle POWER key.

 

CAL UNSECURE is displayed. If CAL SECURE is displayed, then the

 

Counter is already secured against calibration and the remaining steps

 

are not applicable.

2

Press Scale & Offset key until CODE: is displayed.

3

Press the appropriate arrow keys to enter the security code, then

 

press Enter key.

 

2

 

CAL SECURE is displayed.

To Change to a New Security Code

1Press and hold Scale & Offset key, then cycle POWER key.

2Unsecure for calibration by the using the procedure described in the section titled “To Unsecure for Calibration.”

3Press Scale & Offset key until CODE: is displayed.

4Press the appropriate arrow keys to enter the new or customized security code, then press Enter key.

A scrolling message will appear, indicating what the new code is.

For example, the Counter displays the NEW CODE IS 5 if you entered “5” as the new code.

Assembly-Level Service Guide