Chapter 1 Performance Tests

Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity

100 MHz to 1.5 GHz Sensitivity for Option 015 Only

1Set the HP 8663B Synthesized Signal Generator to output a

500 MHz, 60 dBm sine wave and connect the signal to Channel 2 of the Counter as shown in Figure 1-14.

Frequency Standard

HP 8340B(switch set to EXT)

Sweep Oscillator

Output

Input

(rear panel)

 

HP 8663A

 

Synthesized Signal Generator

HP 53181A

50Ω

Counter

Feedthrough

 

Ref. In

 

(rear panel)

Channel 2

Output

N-to-BNC

Connector

Figure 1-14. Option 015/030/050 Channel 2 Frequency Test

2 On the Counter, Press Run key.

1

Assembly-Level Service Guide

1-59

Page 81
Image 81