Chapter 1 Performance Tests

HP 53131A/132A/181A Operational Verification

10 MHz External Timebase Input for the HP 53132A1Connect the equipment as shown in Figure 1-4.

2Set Synthesizer #1 (i.e., the HP 3325B that is connected to the rear-panel Ref In input of the Counter via a 50Ω feedthrough) to output a 10 MHz, 200 mV rms sine wave signal.

3Set Synthesizer #2 (i.e., the HP 3325B that is connected to Channel 1) to output a 2 MHz, 100 mV rms square wave signal.

Verify the Counter displays a reading of approximately 2 MHz and the ExtRef annunciator is lit.

4Mark Pass or Fail in the HP 53131A/132A Performance Test Record on page 1-41,Test 8d.

5Disconnect the test setup.

This completes the Operational Verification for the HP 53132A.