Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
10 MHz External Timebase Input for the HP 53132A
1Connect the equipment as shown in Figure
2Set Synthesizer #1 (i.e., the HP 3325B that is connected to the
3Set Synthesizer #2 (i.e., the HP 3325B that is connected to Channel 1) to output a 2 MHz, 100 mV rms square wave signal.
Verify the Counter displays a reading of approximately 2 MHz and the ExtRef annunciator is lit.
4Mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
5Disconnect the test setup.
This completes the Operational Verification for the HP 53132A.