Chapter 1 Performance Tests

Test 1: Time Interval (HP 53131A/132A Only)

Procedure

1Connect the pulse generator signal to Channel 1 of the Counter as shown in Figure 1-5.

HP 8130A

HP53131A/132Pulse Generator

Counter

Channel 1

Output

Figure 1-5. Time Interval (Channel 1 TO 2) Test Setup

2 Set the Pulse Generator to the following:

PERIOD: 10 μs

WIDTH: 150 ns

High: 1.00 V Low : 0.00 V

Input Mode: Normal

3On the Counter, Press Single/Stop key.

The HP 53131A Counter should display 0.1500 μs ±0.0040 μs. The HP 53132A Counter should display 0.1500 μs ±0.0025 μs.

4Record this value in the appropriate place in the Performance Test Record (Test 1, Line 1 for HP 53131A or Test 1, Line 2 for HP 53132A).

5Disconnect the test setup.