Chapter 1 Performance Tests
Test 1: Time Interval (HP 53131A/132A Only)
Procedure
1Connect the pulse generator signal to Channel 1 of the Counter as shown in Figure
HP 8130A
HP53131A/132Pulse Generator
Counter
Channel 1 | Output |
Figure 1-5. Time Interval (Channel 1 TO 2) Test Setup
2 Set the Pulse Generator to the following:
PERIOD: 10 μs
WIDTH: 150 ns
High: 1.00 V Low : 0.00 V
Input Mode: Normal
3On the Counter, Press Single/Stop key.
The HP 53131A Counter should display 0.1500 μs ±0.0040 μs. The HP 53132A Counter should display 0.1500 μs ±0.0025 μs.
4Record this value in the appropriate place in the Performance Test Record (Test 1, Line 1 for HP 53131A or Test 1, Line 2 for HP 53132A).
5Disconnect the test setup.