Chapter 2 Service
Troubleshooting the Counter
|
| To Run the Individual Self Tests |
1 | Press and hold Recall (Utility) key, then cycle POWER key. | |
|
|
|
NOTE | If power does not cycle or the display does not light, go to the section titled | |
|
| “Power Supply Check” on page |
2 | Press Recall (Utility) key until TEST LOOP: is displayed. | |
3 | Ensure that TEST LOOP:OFF is displayed. | |
|
| Note that the DISP? and KEYPAD? tests are not available if |
|
| TEST LOOP: is ON. |
|
| If TEST LOOP: ON is displayed, press any one of the arrow keys to set |
|
| TEST LOOP to OFF. |
4 | Press Recall (Utility) key until TEST: ALL? is displayed. | |
5 | Press any one of the arrow keys to select the desired self test | |
|
| (for example, the DISP? self test). |
6 | Press Enter key to start the test. | |
|
| After a test is completed, a pass or fail message will be displayed. |
|
|
|
NOTE | After the DISP? (Display) test has cycled through each display segment | |
|
| and annunciator, press the Enter key to stop the test and to display the |
|
| pass or fail message. |
7If you are asked to do something, press Enter key when you have completed the requested task.
After a test is completed, a pass or fail message will be displayed.