Chapter 1 Performance Tests

Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)

Volt Peak 1

1 Connect the equipment as shown in Figure 1-10.

HP 3325B

HP 53131A/132ASynthesizer

Counter

Channel 2

Channel 1

Output

Figure 1-10. Peak Volts Test Setup

2Set the HP 3325B to output a 2 MHz, 2 Vp-p sine wave.

The Counter should display −1.00 ±0.12V for the negative peak of the sine wave. This reading is displayed on the left side of the display.

Record the actual reading in the Performance Test Record (Test 7, Line 1a).

Also, the Counter should display +1.00 ±0.12V for the positive peak of the sine wave. This reading is displayed on the right side of the display.

Record the actual reading in the Performance Test Record (Test 7, Line 1b).