Chapter 1 Performance Tests

Test 1: Time Interval (HP 53131A/132A Only)

Test 1: Time Interval (HP 53131A/132A Only)

This test verifies the Time Interval specifications between Channels 1 and 2.

Equipment

HP 8130A Pulse Generator (or equivalent)

HP 10503A BNC Cable

Counter Setup

1Cycle the POWER key to preset the Counter.

2Press Time & Period key until TI 1 TO 2 is displayed.

3Set CHANNEL 1 input conditions to:

Trigger/Sensitivity, AUTO TRG: OFF

Trigger/Sensitivity, LEVEL: .500V (Enter)

Trigger/Sensitivity, COMMON 1: ON 50Ω/1MΩ, 50Ω

DC/AC, DC

X10 Attenuate, Off 100kHz Filter, Off

4Set CHANNEL 2 input conditions to:

Trigger/Sensitivity, AUTO TRG: OFF

Trigger/Sensitivity, LEVEL: .500V (Enter)

Trigger/Sensitivity, SLOPE: NEG 50Ω/1MΩ, 50Ω

DC/AC, DC

X10 Attenuate, Off 100kHz Filter, Off