Chapter 1 Performance Tests

HP 53131A/132A/181A Operational Verification

HP 53181A

10 MHz

Out

Counter

(rear panel)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Channel 1

Figure 1-1B. Operational Verification Setup for HP 53181A

8Depending on which Counter you are testing, verify the appropriate reading as shown below:

For HP 53131A and HP 53181A, verify a reading of 10.000,000,00 MHz. For the HP 53132A, verify a reading of 10.000,000,000 MHz.

9For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page 1-41, Test 2.

 

9’ For HP 53181A, mark Pass or Fail in the HP 53181A Performance

 

 

Test Record on page 1-71, Test 2.

 

 

The remaining steps apply to the HP 53131A and HP 53132A only

NOTE

 

 

 

since the HP 53181A does not have a standard Channel 2 input.

 

 

For the HP 53181A, proceed to the following “Run Self Test” section.

10Press Freq & Ratio key until FREQUENCY 2 is displayed to select Channel 2 for frequency measurements.

For the HP 53131A, verify a reading of 10.000,000,00 MHz. For the HP 53132A, verify a reading of 10.000,000,000 MHz.

11For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page 1-41, Test 3.

1

Assembly-Level Service Guide

1-7

Page 29
Image 29
HP manual HP 53131A/132A/181A Operational Verification HP 53181A, 1B. Operational Verification Setup for HP 53181A