Chapter 8 HP 53181A Specifications
Instrument Inputs (Continued)
Instrument Inputs (Continued)
Channel 2 Input Specifications3, 4
Frequency Range |
|
|
Opt. 015 | 100 | Mhz to 1.5 Ghz |
Opt. 030 | 100 | MHz to 3 GHz |
Opt. 050 | 200MHz to 5 GHz | |
Opt. 124 | 200 | MHz to 12.4 GHz |
Channel 2 Input Characteristics
Impedance: | 50 Ω |
Coupling: | AC |
VSWR: | < 2.5:1 |
Power Range and Sensitivity (Sinusoid)
Option 015
100 MHz to 1.5 GHz:
Option 030
100 MHz to 2.7 GHz:
2.7 GHz to 3 GHz: | |
Option 050 |
|
200MHz to 5 GHz: |
Option 124
200MHz to 12.4 GHz:
Damage Level: |
|
Option 015, 030 | 5 Vrms |
Option 050, 124 | +25 dBm |
External Arm Input Specifications
Signal Input Range:
TTL compatible |
|
Timing Restrictions: |
|
Pulse Width: | > 50 ns |
Transition Time: | < 250 ns |
> 50 ns | |
Damage Level: | 10 Vrms |
External Arm Input Characteristics
Impedance: | 1 kΩ |
Input Capacitance: | 17 pF |
Start Slope: | Positive or Negative |
Stop Slope: | Positive or Negative |
Notes:
External Arm available for all measurements except Peak Volts.
External Arm is referred to as External Gate for some measurements.
8
3Channel 2 is available as an option.
4When ordered with optional rear terminals, the Channel 2 connector on the front panel for Options 015 or 030 will be removed. There is no degradation in specifications for this input. Option 050 and Option 124 input connectors are available on the front panel only.