Chapter 8 HP 53181A Specifications

Instrument Inputs (Continued)

Instrument Inputs (Continued)

Channel 2 Input Specifications3, 4

Frequency Range

 

 

Opt. 015

100

Mhz to 1.5 Ghz

Opt. 030

100

MHz to 3 GHz

Opt. 050

200MHz to 5 GHz

Opt. 124

200

MHz to 12.4 GHz

Channel 2 Input Characteristics

Impedance:

50 Ω

Coupling:

AC

VSWR:

< 2.5:1

Power Range and Sensitivity (Sinusoid)

Option 015

100 MHz to 1.5 GHz: –27 dBm to +19 dBm

Option 030

100 MHz to 2.7 GHz: –27 dBm to +19 dBm

2.7 GHz to 3 GHz:

–21 dBm to +13 dBm

Option 050

 

200MHz to 5 GHz:

–23 dBm to +13 dBm

Option 124

200MHz to 12.4 GHz: –23 dBm to +13 dBm

Damage Level:

 

Option 015, 030

5 Vrms

Option 050, 124

+25 dBm

External Arm Input Specifications

Signal Input Range:

TTL compatible

 

Timing Restrictions:

 

Pulse Width:

> 50 ns

Transition Time:

< 250 ns

Start-to-Stop Time:

> 50 ns

Damage Level:

10 Vrms

External Arm Input Characteristics

Impedance:

1 kΩ

Input Capacitance:

17 pF

Start Slope:

Positive or Negative

Stop Slope:

Positive or Negative

Notes:

External Arm available for all measurements except Peak Volts.

External Arm is referred to as External Gate for some measurements.