2
Chapter 2 Service
About the HP 53131A/132A Calibration Menu
| The HP 53131A/132A Calibration Menu Tree |
| Turn power off, press and hold Scale & Offset key, then press POWER |
NOTE | |
| key to access this menu. This menu does not exist in early versions |
| (firmware revisions 3317, 3335, and 3402) of the HP 53131A/132A. |
| In the early versions of the Counter, the CAL: menu item resides |
| in the Utility menu, and there is no calibration security |
| capability. See Chapter 6, “Backdating,” in this guide for correct |
| calibration procedures. |
|
|
POWER
| Scale & | ||
| Offset | ||
| On / Stby |
|
|
|
|
|
|
CAL SECURE | CAL UNSECURE | ||
| 1 | CAL: OFFS1? | |
|
| CAL: OFFS2? | |
|
| CAL: GAIN1? | |
|
| CAL: GAIN2? | |
|
| CAL: TI QUIK? | |
|
| CAL: TI FINE? | |
| 2 | CAL: TIMEBAS? | |
|
|
|
|
CODE: 0
CAL COUNT?
HELP: CAL?
HELP: SECURE?
HELP: CODE?
1This menu item appears and calibration is permitted only if calibration is unsecure. Enter in the correct code to change calibration to secure.
2Timebase can be automatically calibrated only if the timebase option is installed.