Chapter 1 Performance Tests

Test 6: Peak Volts, Channel 1 (HP 53181A Only)

Test 6: Peak Volts, Channel 1 (HP 53181A Only)

This test verifies the Peak Volts accuracy specification of the HP 53181A Frequency Counter.

Equipment

HP 3325B Synthesizer

HP 10100C 50Ω Feedthrough

HP 10503A BNC Cable

Counter Setup

1Cycle the POWER key to preset the Counter.

2Press Other Meas key until VOLT PEAKS 1 is displayed.

3Set CHANNEL 1 and CHANNEL 2 inputs to:

50Ω/1MΩ, 50Ω

X10 Attenuate, Off 100kHz Filter, Off