Chapter 1 Performance Tests

Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)

100 kHz to 100 MHz Sensitivity for HP 53181A

1Connect the equipment as shown in Figure 1-12 with the signal generator connected to Channel 1 of the Counter.

 

 

Input

 

 

(rear panel)

 

 

BNC Tee

 

 

HP 8663A

 

50Ω

Synthesized Signal Generator

 

 

 

Feedthrough

 

HP 53181A

Ref. In

 

(rear panel)

 

Counter

 

 

 

Channel 1

Output

N-to-BNC

Connector

Figure 1-12. Channels 1 Frequency Sensitivity Test Setup

2Set the HP 8663A Signal Generator to output a 100 kHz, 35.0 dBm sine wave signal (20.0 dBm for Option 060) to the Counter’s

Channel 1 input.

3In 0.1 dB steps, increase the power level until the Counter displays a stable reading of 100 kHz.

The Counter should display 100 kHz when the signal generator level is ≤ −21.0 dBm (20 mVrms). Counters with Option 060 Rear Terminals should display 100 kHz when the signal generator level is ≤ −9.5 dBm (75 mVrms).

1-52

Assembly-Level Service Guide