Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
100 kHz to 100 MHz Sensitivity for HP 53181A
1Connect the equipment as shown in Figure
|
| Input |
|
| (rear panel) |
|
| BNC Tee |
|
| HP 8663A |
| 50Ω | Synthesized Signal Generator |
|
| |
| Feedthrough |
|
HP 53181A | Ref. In |
|
(rear panel) |
| |
Counter |
| |
|
|
Channel 1 | Output |
Connector
Figure 1-12. Channels 1 Frequency Sensitivity Test Setup
2Set the HP 8663A Signal Generator to output a 100 kHz, −35.0 dBm sine wave signal (−20.0 dBm for Option 060) to the Counter’s
Channel 1 input.
3In 0.1 dB steps, increase the power level until the Counter displays a stable reading of 100 kHz.
The Counter should display 100 kHz when the signal generator level is ≤ −21.0 dBm (20 mVrms). Counters with Option 060 Rear Terminals should display 100 kHz when the signal generator level is ≤ −9.5 dBm (75 mVrms).