Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
200 MHz to 5.0 GHz Sensitivity for HP 53131A/132A (Option 050 only)
1Connect the 8340B Sweep Oscillator to Channel 3 of the counter.
2Change the settings of the Sweep Oscillator to 5.0 GHz, −40 dBm.
3Increase the power level in 1 dB steps until the Counter displays a stable reading of approximately 5.0 GHz.
The Counter should display 5.0 GHz at ≤ −23 dBm.
NOTE | Use a power meter and sensor to ensure the accuracy of the power being |
| input. |
4Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at ≤ −3 dBm.
5Record the actual reading in the Performance Test Record (Test 5, Line 3).
6Disconnect the test setup.