1
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
External Arm Test (HP 53131A/132A Only)
This test verifies the External Arm port of the HP 53131A and HP 53132A Counter is operational.
Equipment
HP 8663A Synthesized Signal Generator
HP 8130A Pulse Generator (or equivalent)
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
HP
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
Other Meas, TOTALIZE 1
Gate & ExtArm, GATE: EXTERNL (ENTER)
3Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: ON
Trigger/Sensitivity, LEVEL: 50 PCT 50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuator, Off 100kHz Filter, Off