Chapter 1 Performance Tests
Test 6: Option 030/050/124 Channel 3 Frequency Accuracy
(HP 53131A/132A Only)
Test 6: Option 030/050/124 Channel 3 Frequency Accuracy (HP 53131A/132A Only)
This test verifies the frequency accuracy of the optional
Equipment
HP 8663A Synthesized Signal Generator (Option 030 only) HP 8340B Sweep Oscillator
HP
HP 10100C 50Ω Feedthrough
HP 11500D
HP
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
Freq & Ratio, FREQUENCY 3
Gate & ExtArm, TIME: TIME 1.000 s
Procedure
1Connect equipment as shown in Figure
(NOTE: For Options 050 and 124, skip steps 2 through 5. Proceed to step 6.)
2Set the HP 8663A Synthesized Signal Generator to output 100 MHz at −17.0 dBm.