Chapter 2 Service

About the HP 53131A/132A Calibration Menu

To View the Calibration Menu and Security Status

The Calibration menu is accessed by holding the Scale & Offset key and cycling the POWER key. The first item in the Calibration menu is a message indicating the calibration security status: CAL SECURE or

CAL UNSECURE.

To Unsecure for Calibration

1Press and hold Scale & Offset key, then cycle POWER key.

CAL SECURE is displayed.

If CAL UNSECURE is displayed, then the Counter is already unsecured for calibration, and the remaining steps are not applicable.

2Press Scale & Offset key until CODE: is displayed.

3Press the appropriate arrow keys to enter the security code “53131”, for example, if 53131 was the security code set at the factory.

4Press Enter key.

CAL UNSECURE is displayed.

NOTE

If “53131” was NOT the correct security code, the Counter would have

 

displayed BAD CODE.

To Initiate the Calibration Routines

1Press and hold Scale & Offset key, then cycle POWER key.

2Unsecure for calibration by performing the preceding procedure.

3Press Scale & Offset key until CAL: OFFS1? is displayed.

2

Assembly-Level Service Guide

2-9