Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
This set of tests verifies the frequency accuracy specifications of the HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator
HP 8340B Sweep Oscillator
HP
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Using the Gate & ExtArm key, set gate time to TIME 1.000 s.
3Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000 V 50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off 100kHz Filter, Off
1