8
Chapter 8 HP 53181A Specifications
Measurement Arming and Processing
Measurement Arming and Processing
Gate Time | Measurement Throughput | |
Auto Mode, or 1 ms to 1000 s |
| 200 Measurements/s (maximum) |
| (See examples in the Programming Guide for ways to optimize | |
| measurement throughput) | |
|
|
|
Measurement Arming |
|
|
Start Measurement: | Free Run, Manual, or External |
|
Stop Measurement: | Continuous, Single, External, or Timed |
|
Arming Modes: |
|
|
(Note: auto arming is the only mode available for the Peak Volts function.) |
| |
Auto Arming: | Measurements are initiated immediately and acquired as fast as possible, using a minimum number of signal | |
| edges. Auto arming offers the highest measurement throughput, though measurement resolution may be | |
| reduced. |
|
Timed Arming: | The duration of the measurement is internally timed to a | |
| This mode should be used when the length of the measurement time must be controlled. | |
Digits Arming: | Measurements are performed to the requested resolution (number of digits) through automatic selection of the | |
| acquisition time. This is the most convenient mode when a specific measurement resolution is desired. | |
External Arming: | An edge on the External Arm Input enables the start of each measurement. Auto Arming, Timed arming modes or | |
| another edge on the External Arm Input may be used to complete the measurement. | |
|
| |
Measurement Statistics |
| |
Available Statistics: | Mean, Minimum, Maximum, Standard Deviation |
|
Number of Measurements: | 2 to 1,000,000. Statistics may be collected on all measurements or on only those which are between the limit | |
| bands. When the Limits function is used in conjunction with Statistics, N (number of measurements) refers to the | |
number of ![]()
N, up to the numerical processing limits of the instrument.
Measurements:Statistics may be collected for all measurements except Peak Volts.
Measurement Limits
Limit Checking:The measurement value is checked against
Display Modes:The measurement result may be displayed as either the traditional numeric value or graphically as an asterisk moving between two vertical bars. These bars define the upper and lower limits, and the asterisk represents the current measurement result relative to these limits.
•The limits annunciator will light on the front panel display.
•The instrument will generate an SRQ if enabled via
•The limits hardware signal provided via the
•If the Analog Display mode is enabled, the asterisk appears outside the vertical bars, which define the upper and lower limits.