Chapter 8 HP 53181A Specifications

Measurement Arming and Processing

Measurement Arming and Processing

Gate Time

Measurement Throughput

Auto Mode, or 1 ms to 1000 s

HP-IB ASCII:

200 Measurements/s (maximum)

 

(See examples in the Programming Guide for ways to optimize

 

measurement throughput)

 

 

 

Measurement Arming

 

 

Start Measurement:

Free Run, Manual, or External

 

Stop Measurement:

Continuous, Single, External, or Timed

 

Arming Modes:

 

 

(Note: auto arming is the only mode available for the Peak Volts function.)

 

Auto Arming:

Measurements are initiated immediately and acquired as fast as possible, using a minimum number of signal

 

edges. Auto arming offers the highest measurement throughput, though measurement resolution may be

 

reduced.

 

Timed Arming:

The duration of the measurement is internally timed to a user-specified value (also known as the “gate time”).

 

This mode should be used when the length of the measurement time must be controlled.

Digits Arming:

Measurements are performed to the requested resolution (number of digits) through automatic selection of the

 

acquisition time. This is the most convenient mode when a specific measurement resolution is desired.

External Arming:

An edge on the External Arm Input enables the start of each measurement. Auto Arming, Timed arming modes or

 

another edge on the External Arm Input may be used to complete the measurement.

 

 

Measurement Statistics

 

Available Statistics:

Mean, Minimum, Maximum, Standard Deviation

 

Number of Measurements:

2 to 1,000,000. Statistics may be collected on all measurements or on only those which are between the limit

 

bands. When the Limits function is used in conjunction with Statistics, N (number of measurements) refers to the

number of in-limit measurements. In general, measurement resolution will improve in proportion to N, up to the numerical processing limits of the instrument.

Measurements:Statistics may be collected for all measurements except Peak Volts.

Measurement Limits

Limit Checking:The measurement value is checked against user-specified limits at the end of each measurement.

Display Modes:The measurement result may be displayed as either the traditional numeric value or graphically as an asterisk moving between two vertical bars. These bars define the upper and lower limits, and the asterisk represents the current measurement result relative to these limits.

Out-of-Limits Indication: The out-of-limits condition can be indicated by any of the following methods:

The limits annunciator will light on the front panel display.

The instrument will generate an SRQ if enabled via HP-IB.

The limits hardware signal provided via the RS-232 connector will go low for the duration of the out-of-limit condition (see the description of this connector under the General Information section of this specifications table).

If the Analog Display mode is enabled, the asterisk appears outside the vertical bars, which define the upper and lower limits.