Chapter 2 Service
The HP 53181A Calibration Procedures
| c. | Press | key once, and then press | key three times. |
|
| The Counter should display CODE: 53000. | ||
| d. | Press | key once, and then press | key once. |
|
| The Counter should display CODE: 53100. | ||
| e. | Press | key once, and then press | key eight times. |
|
| The Counter should display CODE: 53180. | ||
| f. | Press | key once, and then press | key once. |
|
| The Counter should display CODE: 53181. | ||
| g. | Finally, press Enter key. |
| |
|
| CAL UNSECURE is displayed. |
| |
|
| |||
NOTE | If “53181” is NOT the correct security code, the Counter will display | |||
| BAD CODE. |
|
| |
|
|
|
|
|
4Press Scale & Offset key until CAL: OFFS 1 ? is displayed.
5Press Enter key.
A scrolling message on the display will ask you to remove any signals connected to Channel 1. Follow the instruction.
6Press Enter key.
The Counter momentarily displays CALIBRATING, and then it should display OFFS 1 PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
This completes the Offset Calibration procedure. Proceed to the next procedure to calibrate the gain.