Chapter 7 HP 53131A/132A Specifications
Measurement Specifications (Continued)
Measurement Specifications (Continued)
Duty Cycle
Measurement is specified over the full signal range of Channel 1. However, both the positive and negative pulse widths must be greater than 4 ns.
Results Range:0 to 1 (e.g. 50% duty cycle would be displayed as .5)
RMS Resolution: | (tres2 + (2 × Trigger Error2 )) × (1 + Duty Cycle2 ) × Frequency | |
tres | 53131A | 53132A |
750 ps | 300 ps | |
Totalize
Measurement is specified over the full signal range of Channel 1.
Results Range: | 0 to 1015 |
Resolution: | ± 1 count |
Peak Volts
Measurement is specified on Channels 1 and 2 for DC signals; or for AC signals of frequencies between 100 Hz and 30 MHz with
Results Range: | ||
Resolution: | 10 mV |
|
Systematic Uncertainty for AC signals: | 25 mV + 10% of V | |
| for DC signals: | 25 mV + 2% of V |
Use of the input attenuator multiplies all voltage specifications (input range, results range, resolution and systematic uncertainty) by a nominal factor of 10. For example with AC signals, the Systematic Uncertainty becomes: 250 mV + 10% of V.
7