Chapter 2 Service
About the HP 53131A/132A Calibration Menu
| The Calibration menu, illustrated on page |
| • view the calibration security status of the Counter (CAL SECURE or |
| CAL UNSECURE) |
| • initiate the Counter’s calibration routines, which can perform |
| automatic calibration of the voltage offset and gain of the |
| input amplifiers (CAL: OFFS 1 / 2 ?, GAIN 1 / 2?), of the optional |
| timebase (CAL: TIMEBAS?), and of the differences in electrical |
| |
| CAL: TI FINE?) |
| • secure against calibration by entering the security code while the |
| Counter is unsecured (CODE: ) |
| • unsecure for calibration by entering the security code while the |
| Counter is secured (CODE: ) |
| • change the security code by entering a new code while the Counter is |
| unsecured (CODE: ) |
| • view the calibration count, which indicates the number of times that |
| the Counter has been calibrated (CAL COUNT?). The calibration |
| count increments up to a maximum of 32,767 after which it wraps |
| around to 1. A calibration count of 0 indicates that the Counter is |
| completely uncalibrated. (The calibration count is unaffected by |
| interpolator calibration.) |
| • get Help (HELP: ) with any of the following: |
| – how to calibrate (CAL?), |
| – how to secure (SECURE?), and |
| – how to change the code (CODE?) |
|
|
NOTE | A good precaution is to use an |
| example titled “To Read and Store Calibration Data” in Chapter 3 of the |
| Programming Guide) to read and store the calibration factors prior to |
| initiating any calibration. |
|
|
2