1

Chapter 1 Performance Tests

Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP

53181A Only)

Procedure

1Connect equipment as shown in Figure 1-15.

(NOTE: For Option 050/124, skip steps 2 through 7. Proceed to step 8.)

2Set the HP 8663A Synthesized Signal Generator to output 200 MHz at 17.0 dBm.

3On the Counter, press Run key.

Verify the HP 53181A Counter reads 100 MHz ±70 mHz.

4Record the actual reading in the Performance Test Record (Test 5, Line 3).

5Disconnect the HP 8663A output from Channel 2 of the Counter, and connect the HP 8340B Sweep Oscillator output to Channel 2 of the Counter as shown in Figure 1-15.

6Set the HP 8340B to output 3.0 GHz at 17.0 dBm. Verify the HP 53181A Counter reads 3.0 GHz ±2 Hz.

7Record the actual reading in the Performance Test Record. (Test 5, Line 4).

1-66

Assembly-Level Service Guide