Chapter 6 Backdating

The HP 53131A Calibration Procedures

CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)

The HP 53131A Calibration Procedures

To Calibrate the Offset for Channels 1 and 2

1To access the Utility menu, power down, and press and hold the Recall (Utility) key, then press POWER key.

2Press Recall (Utility) key until CAL: OFFS 1 ? is displayed.

3Press Enter key.

The display will ask you to remove any signals connected to Channel 1. Follow the instruction.

4Press Enter key.

The Counter momentarily displays CALIBRATING, and then it should display OFFS 1 PASS.

If the fail message is displayed, refer to the troubleshooting section in this chapter.

5To calibrate Channel 2, press any one of the arrow keys until CAL: OFFS 2 ? is displayed.

6Press Enter key.

The display will ask you to remove any signals connected to Channel 2. Follow the instruction.

7Press Enter key.

The Counter momentarily displays CALIBRATING, and then it should display OFFS 2 PASS.

If the fail message is displayed, refer to the troubleshooting section in this chapter.

6-10

Assembly-Level Service Guide