1
Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
This set of tests verifies frequency sensitivity specifications of the HP 53131A and 53132A Universal Counters.
Equipment
HP 8663A Synthesized Signal Generator
HP
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000 V 50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off 100kHz Filter, Off
3Set CHANNEL 2 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000 V 50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off 100kHz Filter, Off