Chapter 2 Service

About the HP 53131A/132A Calibration Menu

Equipment and Counter Setup: See Figure 2-1.

HP 8130A

In

A B

 

 

HP 53131A/132A

 

Pulse Generator

 

 

Counter

 

 

 

 

 

 

 

 

1

2

3

4

 

 

Output

 

 

 

Start

Stop

 

HP 59992A J06

 

 

 

 

Time Interval Calibrator

Figure 2-1. Calibration Setup for CAL: TI FINE?

Out of HP 8130A, into HP 59992A J06 Calibrator Signal:

PERIOD: 100 ns

WIDTH: 50 ns

High: 0.50 V Low: −0.50 V

Input Mode: Normal

Timebase: Any external timebase you provide is ignored during calibration.

Procedure: From the front-panel calibration menu, one keypress invokes the calibration. You are prompted to press buttons on the HP 59992A J06 Calibrator for each part of a four-part calibration.

Note that CAL: TI FINE? requires the completion of four steps in order.