Chapter 2 Service
About the HP 53131A/132A Calibration Menu
Equipment and Counter Setup: See Figure 2-1.
HP 8130A | In | A B |
|
| HP 53131A/132A |
|
Pulse Generator |
|
| Counter |
| ||
|
|
|
|
| ||
|
| 1 | 2 | 3 | 4 |
|
| Output |
|
|
| Start | Stop |
| HP 59992A J06 |
|
|
|
| |
Time Interval Calibrator
Figure 2-1. Calibration Setup for CAL: TI FINE?
Out of HP 8130A, into HP 59992A J06 Calibrator Signal:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V Low: −0.50 V
Input Mode: Normal
Timebase: Any external timebase you provide is ignored during calibration.
Procedure: From the
Note that CAL: TI FINE? requires the completion of four steps in order.