Chapter 7 HP 53131A/132A Specifications

Measurement Specifications (Continued)

Measurement Specifications (Continued)

HP 53131A—Worst Case RMS Resolution

Automatic or External Arming:

7

Frequency Error (Hz)

 

 

1E+02

1E+00

1E-2

1E-4

1E-6

1E-8

1E-10

10 100

Auto Armed

1ms

10ms

100ms

1s

10s

Gate

Time

1000 10000 100000 1000000 1E+07 1E+08 1E+09 1E+10

Input Frequency (Hz)

Time or Digit Arming:

 

1E+02

 

 

 

 

 

1E+00

 

 

(Hz)

1E-2

 

 

Error

1E-4

 

 

Frequency

 

 

1E-6

 

 

 

 

 

 

1E-8

 

 

 

1E-10

 

 

 

 

 

10

1ms

10ms

100ms

1s

10s

Gate

Time

100 1000 10000 100000 1000000 1E+07 1E+08 1E+09 1E+10

Input Frequency

The preceding graphs do not reflect the effects of trigger error. To place an upper bound on the added effect of this error term, determine the frequency error from the appropriate graph and add a trigger error term as follows:

Automatic or External Arming

 

 

Time or Digit Arming

 

 

 

Frequency Error +

--2----×-----Trigger---------------------Error------------

×

Frequency

Frequency Error +

-----4----×---------2-----×--- Trigger Error -

×

Frequency

or

or

 

Gate Time

 

Period

 

Gate Time ×

Number of Samples

 

Period

7-6

Assembly-Level Service Guide