Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
This set of tests verifies frequency sensitivity specifications of the HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator
HP
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000 V 50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off 100kHz Filter, Off
1