Chapter 2 Service

About the HP 53181A Calibration Menu

• unsecure for calibration by entering the security code while the

 

Counter is secured (CODE: )

 

• change the security code by entering a new code while the Counter is

 

unsecured (CODE: )

 

• view the calibration count, which indicates the number of times that

 

the Counter has been calibrated (CAL COUNT?). The calibration

 

count increments up to a maximum of 32,767 after which it wraps

 

around to 1. A calibration count of 0 indicates that the Counter is

 

completely uncalibrated. (The calibration count is unaffected by

 

interpolator calibration.)

 

• get Help (HELP:) with any of the following:

 

– how to calibrate (CAL?),

 

– how to secure (SECURE?), and

 

– how to change the code (CODE?)

 

 

NOTE

A good precaution is to use an HP-IB program (see the programming

 

example titled “To Read and Store Calibration Data” in Chapter 3 of the

 

Programming Guide) to read and store the calibration factors prior to

 

initiating any calibration.

2-26

Assembly-Level Service Guide