Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
Procedure
1 Set the Pulse Generator to the following:
PERIOD: 20 μs
WIDTH: 5 μs
High: 3.00 V Low : 0.8 V
Trailing/Leading : 1.0 ns
Input Mode: TRIG
2 Connect the equipment as shown in Figure
HP 8130A
Pulse Generator
Output
50Ω |
|
|
|
|
|
|
|
|
| HP 8663A | ||||
Feedthrough |
|
|
| Synthesized Signal Generator | ||||||||||
Ext Arm |
|
|
|
|
|
|
| |||||||
(rear panel) |
|
|
|
|
|
|
| |||||||
Counter |
|
|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Channel 1 | Output |
Connector
Figure 1-3. External Arm Test Setup (HP 53131A/132A Only)
1