HP 181A manual Procedure, HP 8130A Pulse Generator, HP 8663A, Synthesized Signal Generator

Models: 181A

1 254
Download 254 pages 46.13 Kb
Page 33
Image 33

Chapter 1 Performance Tests

HP 53131A/132A/181A Operational Verification

Procedure

1 Set the Pulse Generator to the following:

PERIOD: 20 μs

WIDTH: 5 μs

High: 3.00 V Low : 0.8 V

Trailing/Leading : 1.0 ns

Input Mode: TRIG

2 Connect the equipment as shown in Figure 1-3.

HP 8130A

Pulse Generator

Output

50Ω

 

 

 

 

 

 

 

 

 

HP 8663A

Feedthrough

 

 

 

Synthesized Signal Generator

Ext Arm

 

 

 

 

 

 

 

(rear panel)

 

 

 

 

 

 

 

Counter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Channel 1

Output

N-to-BNC

Connector

Figure 1-3. External Arm Test Setup (HP 53131A/132A Only)

1

Assembly-Level Service Guide

1-11

Page 33
Image 33
HP 181A manual Procedure, HP 8130A Pulse Generator, HP 8663A, Synthesized Signal Generator