Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
2.8to 3.0 GHz Sensitivity for Option 030 Only
1Disconnect the HP 8663A Synthesized Signal Generator from the Counter (see Figure
2Change the settings of the Sweep Oscillator to 3.0 GHz, −40 dBm.
3Increase the power level in 1 dB steps until the Counter displays
3.0GHz.
The Counter should display 3.0 GHz at ≤ −21.0 dBm.
4Sweep the frequencies from 3.0 GHz to 2.8 GHz.
The Counter should read frequencies from 3.0 to 2.8 GHz at ≤ −21.0 dBm.
5Record the actual reading in the Performance Test Record (Test 4, Line 3).
6Disconnect the test setup.
200 MHz to 5.0 GHz Sensitivity for Option 050 only
1Connect the 8340B Sweep Oscillator to Channel 2 of the counter.
2Change the settings of the Sweep Oscillator to 5.0 GHz, −30 dBm.
3Increase the power level in 1 dB steps until the Counter displays a stable reading of approximately 5.0 GHz.
The Counter should display 5.0 GHz at ≤ −23 dBm.
NOTE | Use a power meter and sensor to ensure the accuracy of the power being |
| input. |
|
|
4Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at ≤ −23 dBm.
5Record the actual reading in the Performance Test Record (Test 4, Line 4).
1