Chapter 1 Performance Tests
Test 6: Peak Volts, Channel 1 (HP 53181A Only)
Procedure
1 Connect the equipment as shown in Figure
HP 3325B
HP 53181ASynthesizer
Counter
1
Channel 1 | Output |
Figure 1-16. Peak Volts Test Setup
2Set the HP 3325B to output a 2 MHz, 2
The Counter should display −1.00 ±0.12V for the negative peak of the sine wave. This reading is displayed on the left side of the display.
Record the actual readings in the Performance Test Record (Test 6, Line a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the sine wave. This reading is displayed on the right side of the display.
Record the actual readings in the Performance Test Record (Test 6, Line b).
3Disconnect the test setup.