Chapter 1 Performance Tests

Test 6: Peak Volts, Channel 1 (HP 53181A Only)

Procedure

1 Connect the equipment as shown in Figure 1-16.

HP 3325B

HP 53181ASynthesizer

Counter

1

Channel 1

Output

Figure 1-16. Peak Volts Test Setup

2Set the HP 3325B to output a 2 MHz, 2 Vp-p sine wave, and connect the signal to Channel 1 of the Counter.

The Counter should display −1.00 ±0.12V for the negative peak of the sine wave. This reading is displayed on the left side of the display.

Record the actual readings in the Performance Test Record (Test 6, Line a).

Also, the Counter should display +1.00 ±0.12V for the positive peak of the sine wave. This reading is displayed on the right side of the display.

Record the actual readings in the Performance Test Record (Test 6, Line b).

3Disconnect the test setup.

Assembly-Level Service Guide

1-69