1
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
5’ For HP 53181A, mark Pass or Fail in the HP 53181A Performance Test Record on page
5 MHz External Timebase Input (HP 53131A and HP 53181A Only)
1Change the frequency of Synthesizer # 1, which is connected to the Counter’s
2On the Counter, press Run key.
Verify the Counter displays a reading of approximately 2 MHz and the ExtRef annunciator is lit.
3For HP 53131A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
3’ For HP 53181A, mark Pass or Fail in the HP 53181A Performance Test Record on page
10 MHz External Timebase Input (HP 53131A and HP 53181A Only)
1Change the frequency of Synthesizer #1, which is connected to the Counter’s
2On the Counter, press Run key.
Verify the Counter displays a reading of approximately 2 MHz and the ExtRef annunciator is lit.
3For HP 53131A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
3’ For HP 53181A, mark Pass or Fail in the HP 53181A Performance Test Record on page
4Disconnect the test setup.
This completes the Operational Verification for the HP 53131A and HP 53181A.