HP manual HP 53181A Performance Test Record Page 2

Models: 181A

1 254
Download 254 pages 46.13 Kb
Page 94
Image 94

Chapter 1 Performance Tests

HP 53181A Performance Test Record (Page 2 of 4)

1

HP 53181A Performance Test Record (Page 2 of 4)

Complete Performance Tests

Test

 

 

 

 

 

 

 

 

 

 

Number

 

Test Description

 

Minimum

 

Actual Reading

Maximum

 

 

 

 

 

 

 

 

 

 

 

1

 

Trigger Level:

 

 

 

 

 

 

 

 

 

 

Channel 1

 

 

−15 mV

 

 

1. __________

+15 mV

 

 

 

 

 

−15 mV

 

 

2. __________

+15 mV

 

 

 

 

 

 

 

 

 

 

Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)

 

 

 

 

Standard (Front Panel Terminals)

 

Actual Reading

 

 

Specification

 

 

 

 

 

 

 

 

 

100 kHz–100 MHz, Channel 1

 

 

1. __________

 

 

≤ −21 dBm (20 mVrms)

 

 

 

 

 

 

 

 

 

100 MHz–200 MHz, Channel 1

 

 

2. __________

 

 

≤ −17.5 dBm (30 mVrms)

 

 

 

 

 

 

 

 

 

200 MHz–225 MHz, Channel 1

 

 

3. __________

 

 

≤ −15.0 dBm (40 mVrms)

 

 

 

 

 

 

 

 

 

 

Option 060 Rear Terminals

 

 

Actual Reading

 

 

Specification

 

 

 

 

 

 

 

 

 

100 kHz–225 MHz, Channel 1

 

 

1. __________

 

 

≤ −9.5 dBm (75 mVrms)

 

 

 

 

 

 

 

 

 

 

 

Test

 

 

 

 

 

 

Actual

 

 

Number

Test Description

 

Minimum

 

Reading

Maximum

 

 

 

 

 

 

 

 

 

 

3

Channel 1

 

 

 

 

 

 

 

 

 

Frequency

 

 

 

 

 

 

 

 

 

Accuracy

 

 

 

 

 

 

 

 

 

(HP 53181A Only)

 

 

 

 

 

 

 

 

 

1 MHz Test

999,999.999 Hz

1. __________

1, 000, 000.001 Hz

 

100 MHz Test

99.999,999,93 MHz

2. __________

100.000, 000,07 MHz

 

200 MHz Test

199.999,999,87 MHz

3. __________

200.000,000,13 MHz

 

225 MHz Test

224.999,999,85 MHz

4. __________

225.000,000,15 MHz

 

 

 

 

 

 

 

 

 

 

 

1-72

Assembly-Level Service Guide

Page 94
Image 94