Chapter 1 Performance Tests

Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)

 

Volt Peak 2

 

1

Remove the signal from Channel 1 of the Counter and connect it

 

 

to Channel 2.

 

2

Press Other Meas key until VOLT PEAK 2 is displayed.

1

 

 

The Counter should display -1.00 ±0.12V for the negative peak of the sine

 

 

wave. This reading is displayed on the left side of the display.

 

 

 

 

Record the actual reading in the Performance Test Record

 

(Test 7, Line 2a).

 

Also, the Counter should display +1.00 ±0.12V for the positive peak of the

 

sine wave. This reading is displayed on the right side of the display.

 

Record the actual reading in the Performance Test Record

 

(Test 7, Line 2b).

3

Disconnect the test setup.

Assembly-Level Service Guide

1-39