Chapter 2 Service
The HP 53131A/132A Calibration Procedures
5Press Enter key.
The Counter displays CALIBRATING, and then it should display
TI CAL PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
6Disconnect the calibration setup.
This completes the TI QUIK? calibration procedures.
Depending on which timebase the Counter contains, proceed to the procedure “To Calibrate the Standard Timebase” (page
CAL: TI FINE? Calibration
1Connect the HP 8130A Pulse Generator, HP 59992A J06 Time Interval Calibrator, and HP 53131A/123A Counter as shown in Figure
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 59992A J06 Time Interval Calibrator (or equivalent) HP 10503A BNC Cable (3)
2
HP 8130A
Pulse Generator
Output
In A B
1 ![]()
2 ![]()
3 ![]()
4 ![]()
![]()
HP 59992A J06
Time Interval Calibrator
HP 53131A/132A Counter
Start Stop