Chapter 1 Performance Tests
Test 2: Trigger Level (HP 53131A/132A Only)
Procedure
1Connect the output of the HP 3325B to Channel 1 of the Counter as shown in Figure
HP 3325B
HP 53131A/132A Synthesizer
Counter
Channel 1 | Channel 2 | Output |
Figure 1-6. Trigger Level Test Setup
2Set the HP 3325B to output a 1 MHz, 80
3Set the HP 3325B DC OFFSET to −60 mV.
4On the Counter, press Run key.
5Now, increment the dc offset on the HP 3325B by +1 mV until the Counter’s Gate annunciator flashes and continue incrementing until the Counter displays a stable reading of approximately
1 MHz.
Observe the offset value on the display of the HP 3325B. Record the dc offset value__________ mV.
6Add the upper peak voltage (40 mV) of the 80 mV
This is the upper hysteresis level.
Record the upper hysteresis level__________ mV.