Chapter 1 Performance Tests

Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)

100 kHz to 100 MHz Sensitivity for HP 53131A/132A

1Connect the timebase output of the unit with the better timebase to the timebase reference input of the other unit as shown in Figure 1-7.

For this test, make sure you always use the unit (HP 53131A/132A or HP 8663A) that contains the better 10 MHz timebase as the output source as shown in Figure 1-7.

1

50Ω Feedthrough

HP 53131A/132A Ref. In

(rear panel)

Counter

HP 8663A

Synthesized

Signal Generator

Output (rear panel)

N-to-BNC Connector

Channel 1

Output

OR

HP 53131A/132A Counter

Channel 1

HP 8663A

 

Synthesized

Input

Signal Generator

(rear panel)

50Ω

 

Feedthrough

 

10 MHz

 

Output

 

(rear panel)

N-to-BNC

 

 

Connector

 

Output

Figure 1-7. Channels 1 and 2 Frequency Sensitivity Test Setup

Assembly-Level Service Guide

1-23

Page 45
Image 45
HP 181A KHz to 100 MHz Sensitivity for HP 53131A/132A, HP 53131A/132A Ref, Counter HP 8663A Synthesized Signal Generator