Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
100 kHz to 100 MHz Sensitivity for HP 53131A/132A
1Connect the timebase output of the unit with the better timebase to the timebase reference input of the other unit as shown in Figure
For this test, make sure you always use the unit (HP 53131A/132A or HP 8663A) that contains the better 10 MHz timebase as the output source as shown in Figure
1
50Ω Feedthrough
HP 53131A/132A Ref. In
(rear panel)
Counter
HP 8663A
Synthesized
Signal Generator
Output (rear panel)
Channel 1 | Output |
OR
HP 53131A/132A Counter
Channel 1
HP 8663A |
|
Synthesized | Input |
Signal Generator | (rear panel) |
50Ω |
|
Feedthrough |
|
10 MHz |
|
Output |
|
(rear panel) | |
| |
| Connector |
| Output |