Chapter 1 Performance Tests

Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)

4

Record the actual reading in the Performance Test Record

 

 

(Test 3, Line 1).

 

5

Change the signal generator frequency to 100 MHz.

1

 

 

 

Verify the Counter reads 100 MHz ±70 mHz.

 

6

Record the actual reading in the Performance Test Record

 

 

(Test 3, Line 2).

 

 

 

7

Change the signal generator frequency to 200 MHz.

 

Verify the Counter reads 200 MHz ±130 mHz.

8

Record the actual reading in the Performance Test Record

 

(Test 3, Line 3).

9

Change the signal generator frequency to 225 MHz.

 

Verify the Counter reads 225 MHz ±150 mHz.

10

Record the actual reading in the Performance Test Record

 

(Test 3, Line 4).

11

Disconnect the test setup.

Assembly-Level Service Guide