Chapter 7 HP 53131A/132A Specifications
Measurement Arming and Processing (Continued)
Measurement Arming and Processing (Continued)
Measurement Statistics
Available Statistics: Mean, Minimum, Maximum, Standard Deviation
Number of Measurements: 2 to 1,000,000. Statistics may be collected on all measurements or on only those which are between the limit bands. When the Limits function is used in conjunction with Statistics, N (number of measurements) refers to the number of ![]()
N, up to the numerical processing limits of the instrument.
Measurements:Statistics may be collected for all measurements except Peak Volts and Totalize.
Measurement Limits
Limit Checking:The measurement value is checked against
Display Modes:The measurement result may be displayed as either the traditional numeric value or graphically as an asterisk moving between two vertical bars. These bars define the upper and lower limits, and the asterisk represents the current measurement result relative to these limits.
•The limits annunciator will light on the front panel display.
•The instrument will generate an SRQ if enabled via
•The limits hardware signal provided via the
•If the Analog Display mode is enabled, the asterisk appears outside the vertical bars, which define the upper and lower limits.