Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
1 MHz External Timebase Input (HP 53131A and HP 53181A Only)
1 Connect the equipment as shown in Figure
HP 3325B
Synthesizer #1
50Ω Feedthrough
Ref. In
(rear panel)
Output
HP 3325B
Synthesizer #2
Counter
Channel 1 | Output |
Figure 1-4. External Timebase Test Setup
2Cycle the POWER key to preset the Counter.
3Set Synthesizer #1 (i.e., the HP 3325B connected to the
1 MHz, 200 mV rms sine wave signal.
4Set Synthesizer #2 (i.e., the HP 3325B connected to Channel 1) to output a 2 MHz, 100 mV rms square wave signal.
Verify the Counter displays a reading of approximately 2 MHz and the ExtRef annunciator is lit.
5For HP 53131A, mark Pass or Fail in the HP 53131A/132A Performance Test Record on page
1