Chapter 1 Performance Tests

Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity

Test 4: Option 015/030/050/124 Channel 2

Frequency Sensitivity

1

This test verifies the frequency range and sensitivity of the optional

1.5-GHz (Option 015), 3-GHz (Option 030), 5.0-GHz (Option 050), or

12.4-GHz (Option 124) Channel 2 for the HP 53181A Frequency Counter.

Equipment

HP 8663A Synthesized Signal Generator (Not needed for Option 050/124)

HP 8340B Sweep Oscillator

HP 1250-0780 Type N-to-BNC Connector (Not needed for Option 050/124)

HP 10503A BNC Cable (2)

HP 10100C 50Ω Feedthrough

Counter Setup

1Cycle the POWER key to preset the Counter.

2Set the Counter as follows: Freq & Ratio, FREQUENCY 2

1-58

Assembly-Level Service Guide