Chapter 1 Performance Tests

Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)

Procedure

1Connect the equipment as shown in Figure 1-13with the signal generator connected to Channel 1 of the Counter.

1

Frequency Standard

HP 8340B(switch set to EXT)

Sweep Oscillator

Output

Input

(rear panel)

HP 8663A

Synthesized Signal Generator

HP 53181A Counter

50Ω

Feedthrough

Ref. In

(rear panel)

Channel 1

Output

N-to-BNC

Connector

Figure 1-13. Channels 1 Frequency Accuracy Test Setup

2Set the HP 8663A Signal Generator to output 1 MHz at 7 dBm.

3On the Counter, press Run key.

Verify the Counter reads 1 MHz ±1 mHz.

1-56

Assembly-Level Service Guide