Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
Procedure
1Connect the equipment as shown in Figure
HP 53181A Counter
Frequency Standard
HP 8340B(switch set to EXT)
Sweep Oscillator
Output
Input
(rear panel)
HP 8663A
Synthesized Signal Generator
50Ω
![]()
![]()
Feedthrough
Ref. In
(rear panel)
Channel 1 | Output |
Connector
Figure 1-13. Channels 1 Frequency Accuracy Test Setup
2Set the HP 8663A Signal Generator to output 1 MHz at −7 dBm.
3On the Counter, press Run key.
Verify the Counter reads 1 MHz ±1 mHz.