Chapter 6 Backdating

The HP 53131A Calibration Procedures

3Connect the HP 8130A Pulse Generator output to Channel 1 of the Counter as shown in Figure 2-3.

CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)

Equipment

HP 8130A Pulse Generator (or equivalent) HP 10503A BNC Cable

HP 8130A

HP53131APulse Generator

Counter

6

Channel 1Output

Figure 2-3. TI Calibration Setup

 

4 Set the HP 8130A as follows:

 

 

PERIOD: 100 ns

 

 

WIDTH: 50 ns

 

 

High: 0.50 VLow : −0.50 V

 

 

Input Mode: Normal

 

 

 

NOTE

 

Any external timebase connected to the Counter is ignored during this

 

 

calibration.

 

 

 

5Press Enter key.

The Counter displays CALIBRATING, and then it should display

TI CAL PASS.

If the fail message is displayed, refer to the troubleshooting section in this chapter.

Assembly-Level Service Guide

6-13