Chapter 6 Backdating
The HP 53131A Calibration Procedures
3Connect the HP 8130A Pulse Generator output to Channel 1 of the Counter as shown in Figure
CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)
Equipment
HP 8130A Pulse Generator (or equivalent) HP 10503A BNC Cable
HP 8130A
HP53131APulse Generator
Counter
6
Channel 1Output
Figure 2-3. TI Calibration Setup
| 4 Set the HP 8130A as follows: | |
|
| PERIOD: 100 ns |
|
| WIDTH: 50 ns |
|
| High: 0.50 VLow : −0.50 V |
|
| Input Mode: Normal |
|
|
|
NOTE |
| Any external timebase connected to the Counter is ignored during this |
|
| calibration. |
|
|
|
5Press Enter key.
The Counter displays CALIBRATING, and then it should display
TI CAL PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.