HP 181A manual Location of the Rubber Foot

Models: 181A

1 254
Download 254 pages 46.13 Kb
Page 164
Image 164

3

Chapter 3 Replacing Assemblies

To Remove A4 AC Power Supply Assembly

3Remove the two TORX screws (H2) on the rear panel using the TORX 10 screwdriver as shown in Figure 3-9.

4Lift up A4 AC Power Supply Assembly and disconnect its cable from J18 on A1 Motherboard Assembly to remove the power supply assembly as shown in Figure 3-9.

NOTE

When replacing the power supply, order the rubber foot (0403-0424).

 

Figure 3-10 shows where the rubber foot is located on the defective

 

assembly.

 

 

Rubber foot

A4 AC Power

Supply Assembly

Figure 3-10. Location of the Rubber Foot

5If the Counter contains the Option 002 DC Power Input Assembly, disconnect A4 AC Power Supply’s other cable from Option 002 as shown in Figure 3-11.

3-16

Assembly-Level Service Guide

Page 164
Image 164
HP 181A manual Location of the Rubber Foot